Calibration
The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to the use instructions, periodic readjustment is not required. If at any time you question the accuracy of temperature measure- ments, please contact Braun Infoline (see page 2).
Manufacturing date is given by the LOT number located in the battery compartment. The first number after LOT represents the last digit of the year of manufacture.
The last two digits give the week of manufacture. An example: LOT 503 – this product has been manufactured in the week 03 of the year 2005.
Product specifications
Displayed temperature |
|
range: | 34 °C – 42.2 °C |
Operating ambient | (93.2 °F – 108 °F) |
| |
temperature range: | 10 °C – 40 °C |
| (50 °F – 104 °F) |
Display resolution: | 0.1 °C or °F |
Accuracy for displayed |
|
temperature range: | ± 0.2 °C |
| |
| ± 0.3 °C (outside this |
| temperature range) |
Clinical repeatability: | ± 0.25 °C (± 0.45 °F) |
Humidity: | ≤ 95% |
Battery life: | 2 years / |
| 1000 measurements |
If device is not used within specified temperature and humidity ranges the technical accuracy can not be ensured.
Type BF equipment
Subject to change without notice.
This appliance conforms to the following standards:
DIN EN
Part 1: General requirements for safety
DIN EN
Part 5: Performance of infrared ear thermometers (with maximum device)
This product conforms to the provisions of the EC directive 93/42/EEC (Medical Device Directive).
0297
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of EMC requirements please contact the Braun Infoline (see page 2).
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
Please do not dispose of the product in the household waste at the end of its useful life. Disposal can take place at a Braun Service Centre or at appropriate collection points provided in your country.
7