
17 
4.4 TRIGGERED MODES OF OPERATION 
The  pulse stream  can  be  programmed  to  run  continuously,  o ut put  a  single function  when 
triggered, or output continuously as long as  the gate (trigger in) is high (i.e.,  free run, trigger, or 
gated repetition.).  All three trigger modes are described below.  The RUN control bit must be set 
for any  mode of o peratio n to pr oduce a  pulse st ream. 
4.4.1 Free 
When the FRE and RUN bits are programmed high, a continuous stream of pulses are generated 
at the out put driver.  Bo th normal and special modes o f operation function in the free run  mode.  
The  triggered  modes  operate  only  with  t he  FRE  bit  set  to  0.    Each  pulse cycle will be 
accompanied with a Trigger Out to indicate the beginning of the cycle. 
4.4.2 Triggered Mode 
When the  FRE bit  is reset ( low) and  the RUN bit  is set  (high), the  output  pulse str eam function  is 
dependent on the TMD contro l bit.  If TMD is reset (low) for tr iggered mode, an output pulse or 
a double pulse is generated o n each rising edge of the trigger  pulse.  Any triggers occurring prior 
to the end of the pulse  repetition interval are ignored, even if the  pulse has finished.  This allows 
the pulse repetit ion inter val to  be used  to ho ld off re trigger ing. 
4.4.3 Gated Mode 
When the  FRE bit  is reset ( low) and  the RUN bit  is set  (high), the  output  pulse str eam function  is 
dependent on the TMD control bit.  If TMD  is set (high) for gated mode, the output pulse stream 
performs like free run as long  as the input gate  (trigger in) is high.  When the gate goes  low, the 
current  pulse stream  completes  then stops.    For  square wave  modes,  the level  remains at  the 
current level. 
4.5 SPECIAL MODES OF OPERATION 
Along  with  the  normal  modes  of o peration,  three  special  free  running  square  wave  modes  are 
provided.   These include a  frequency programmable square wave  and fixed frequency  20 MHz 
and 40  MHz square  waves.   These modes pro vide a rapid  method of pro ducing a square wave 
output.   The  square  wave generation logic  produces an  output level change  on each pulse 
repetition interval.   Note  that  this results  in a  square wave  with a  period equal  to  twice the 
programmed PRI.  The  pulse width and delay time registers are  disabled in this mode.  In the 2 0 
and 40 MHz modes the  PRI, Pulse Width, and Delay time registers are  disabled.  Additionally, at 
these high  frequencies the output has limited output voltage capability; however, 5 Vpp can still 
be sustained.