Application Programs
Counting
Purpose
With the Counting program you can determine the number of parts that each have approximately equal weight.
Available Features
–Store the current weight value to have it loaded as the preset reference sample quantity next time you initialize the Counting application
–The reference sample quantity can be changed in the operating menu: See “Configuring the Scale”
–The average piece weight is automatically output via the optional data interface port after initialization, if the menu code for “Printout with data ID codes” is set
–Press [FUNCTION] to toggle the display between piece count and weight
Factory Settings
Reference sample quantity: 10 (menu code 3. 3. 2)
Preparation
•Configure the Counting application in the
operating menu:
See “Configuring the Scale”
Menu code 2. 1. 4 Counting
•Reference sample quantity:
Menu code 3. 3. 1 | 5 pcs |
Menu code 3. 3. 2 | 10 pcs |
Menu code 3. 3. 3 | 20 pcs |
Menu code 3. 3. 4 | 50 pcs |
Menu code 3. 3. 5 | 100 pcs |
See also “Configuring the Scale”
Example
Determine an unknown piece count; weigh the preset reference sample quantity
Settings (changes in the factory settings required for this example):
Menu: Application program: Counting (2 .1. 4) Reference sample quantity: 20 pcs (3. 3. 3)
Step |
| Key (or instruction) Display | ||||
1. | Turn on the scale | [ON/OFF] |
|
|
| |
2. | Tare the scale | t |
| 0.000 c t | ||
3. | Display the reference | [FUNCTION] |
| ref | 20 (briefly) | |
| sample | quantity | > 2 sec. |
|
|
|
| (here: | 20 pcs) |
|
|
|
|
4. | Place the reference |
| + | 16.000 c t | ||
| sample | quantity |
|
|
|
|
| (20 pcs) on the |
|
|
|
| |
| scale (here: 16 ct) |
|
|
|
| |
5. | Start the application; | [FUNCTION] | + |
| 20 p c s | |
| if the print format is set |
|
|
|
| |
| to include data ID codes, |
|
|
|
| |
| the following is printed |
| wRef | + 0.8000 ct | ||
6. | Weigh | uncounted parts |
| + |
| 174 p c s |
| (here: 174 pcs) |
|
|
|
| |
7. | Display weight | [FUNCTION] | +139.200 c t | |||
8. | Display | quantity | [FUNCTION] | + |
| 174 p c s |
9. | Unload the scale |
|
|
| 0 p c s | |
10. | Delete the reference | [CAL/CF] > 2 sec. |
|
|
| |
| value |
|
|
|
|
11.Repeat the procedure starting from Step 6, if desired.
Weighing in Percent
Purpose
This application program allows you to obtain weight readouts in percent which are in proportion to a reference weight.
Available Features
–Store the current weight value to have it loaded as the preset reference percentage next time you initialize the Weighing in Percent application
–The reference percentage can be changed
in the operating menu: See “Configuring the Scale”
–The reference percentage is automatically output via the optional data interface port after initialization, if the menu code for “Printout with data ID codes” is set
–Press [FUNCTION] to toggle the display between percentage and weight
Factory Settings
Reference percentage: 10 (menu code 3. 3. 2)
4