The LTL2000S(Q) is factory calibrated. Nevertheless start important measurements with a calibration. Study the Retroreflectometer status and error messages if any. See also Section 4

– Maintenance (page 25).

When the battery voltage becomes low a warning is shown in the display (W). It is still possi- ble to take measurements but when the voltage drops further, readings can be erroneous.

If in doubt whether the readings are correct make a control measurement on the calibration unit.

Note

Keep light trap, optical-protection window and ceramics on calibration unit clean.

Keep battery fully charged. A well charged battery is more resistant to ageing and damage by freezing

DELTA

LTL2000S(Q) Retroreflectometer