Command set
Command
Parameter
Response
Meaning
OTn
PV
QCn
QEn
RCn
REn
RZ
SD
SNn
TIn
| Rl Status Code | 0 00000000 | Rl status code | |
| 2002 Aug 12 13:53:19 | Time for Rl measurement | ||
| LTL2000SQ v1.0 s/n:n | Instrument Id | ||
None | Off Timer = 300 sec. | Automatic turn off time when not in use. Range 60 to | ||
59 | Off Func: Off |
| 600 sec. Disable for time < 60 sec | |
180 | Off Timer: = 180 sec. |
| ||
None | V_PMT = 644 Volt | Voltage on PMT | ||
None | QdF: 0.1366 |
| Query Qd calc factor | |
123 | Cal Qd? [Y/N] |
| Calibrate Qd to 123 Y/N response to start | |
| Cal Qd! |
| Calibrating Qd | |
| Qd Factor = 0.7564 | Calculated calc factor | ||
| Qd Normal = 123 | New active Qd normal | ||
None | Qd | On |
| Query Qd function |
F | Qd | Off |
| Set Qd function Off |
T | Qd | On |
| Set Qd function On |
None | RlF: 2.038 |
| Query Rl Calc factor | |
123 | Cal Rl [Y/N] |
| Calibrate Rl to 123 Y/N response to start | |
| Cal Rl |
| Calibrating Rl | |
| Rl Factor = 2.032 | Calculated calc factor | ||
| Rl Normal = 123 | New active Rl normal | ||
None | Rl | On |
| Querry Rl function |
F | Rl | Off |
| Set Rl function Off |
T | Rl | On |
| Set Rl function On |
None | Zero Adjust? [Y/N] | Perform Rl zero adjust with Y/N confirmation | ||
| Zero Rl |
| Measuring zero signal | |
| *** Zero Reading *** | Result from zero reading | ||
| Lamp On: | 12 | Signal with lamp on | |
| Lamp Off: | 10 | Signal with lamp off | |
| New Rl_Sy_Leak: 2 | New SyLeak signal | ||
None | Rl Status Code | 0 00000000 | Status code and bit pattern | |
| Qd Status Code: 0 00000000 | if status not zero then the status code is decoded | ||
None | Seq ID: DELTA | Querry Sequence number | ||
| Meas in Seq: 3 |
| Measurements with this ID | |
TEST2 | Seq ID: TEST2 |
| Change ID to TEST2 | |
| Meas in Seq: 0 |
| No measurements with this ID | |
None | 2002 Aug 12 13:59:37 | Real time clock Date and Time | ||
10 02 25 | 2002 Aug 12 10:02:25 | New Time (Date unchanged) |
DELTA | LTL2000S(Q) Retroreflectometer | 34 |