Delta LTL2000S manual Display shows Step

Models: LTL2000S

1 45
Download 45 pages 57.18 Kb
Page 11
Image 11

The display shows: Step 1

Mount Light Trap when Ready

Step 2

Zeroing Please Wait

Step 3

Zero OK Zero: 0.05%

Step 4

Mount Rl Normal when ready

Step 5

Calibrating Please Wait

Step 6

Rl Normal 150

= OK to Edit Wait

Step 7

Rl Normal 150 = when OK

Step 8

Calibration Done Remove Normal

To verify the calibration do a Rl measurement on the normal before removing the normal in step 8.

To document the calibration press the key to print.

In the example above the Rl function is now calibrated to a Rl value of 150.

If an error is detected during the calibration procedure then the display will indicate this.

Note:

Avoid calibration in direct sunlight

Store the reference calibration unit in a dry and clean environment.