Eagle Home Products Eagle Series TWA/STEL Screen, Battery Voltage Screen, Date/Time Screen

Models: Eagle Series

1 80
Download 80 pages 15.42 Kb
Page 27
Image 27

TWA/STEL Screen

The TWA/STEL screen displays the time-weighted average (TWA) and the short-term exposure limit (STEL) readings for toxic gases only.

The TWA reading is the average reading during the last 8 hours. If 8 hours have not elapsed since the last time the TWA/STEL reading was cleared, the average is still calculated over 8 hours. The missing time is assigned a 0 value for readings.

The STEL reading is the average reading during the last 15 minutes.

Battery Voltage Screen

The Battery Voltage screen displays the minimum operating voltage and present battery voltage. New alkaline batteries typically measure

6.0V; fully-charged Ni-Cd batteries typically measure 5.2 V. This screen also displays when you turn on the Eagle.

NOTE: The remaining screens only display if your Eagle includes the data logging option. If your Eagle does not include the data logging option, press the DISP/ADJ button to return to the normal screen.

Date/Time Screen

The date/time screen displays the current date and time. You can set the date and time in Calibration or Setup mode (see page 38).

Eagle Instruction Manual

Display Mode • 21

Page 27
Image 27
Eagle Home Products Eagle Series instruction manual TWA/STEL Screen, Battery Voltage Screen, Date/Time Screen