Updating Channel Settings

This procedure describes how to update channel settings for the combustible gas, oxygen, and toxic gas channels.

CAUTION:Verify that the correct sensor is installed before you update a channel’s settings.

Updating combustible gas channel settings

This section describes how to update the target gas label, set a custom gas label, and update the fullscale PPM setting for the combustible gas channel.

Updating the target gas label

1.From the main menu, select the GAS COMBINATIONS menu option.

2.Use the AIR/or SHIFT/button to place the prompt next to the combustible gas channel (in this example CH4).

3.Press the POWER/ENTER button. The combustible gas target gas label flashes. This indicates that this setting can now be updated.

4.Press the AIR/or SHIFT/buttons to display the available combustible gas target gas labels (CH4, HEX, H2, ***, and NOT USED).

NOTE: Select the HEX or *** setting for Methane Elimination

(see “Appendix B: Methane Elimination” on page 52 for more information.)

5.Press the POWER/ENTER button to enter the new target gas label.

A screen displays that shows the fullscale PPM setting, which corresponds to 100% LEL, and display increments for the target gas label you selected. If you select *** as the gas label, you must update the fullscale PPM setting to correspond to 100% LEL for the target gas.

Eagle Instruction Manual

Setup Mode • 25