Emerson 410VP Setup And Calibration, Troubleshooting, Problem, Probable Cause, Solution

Models: 410VP

1 4
Download 4 pages 18.13 Kb
Page 3
Image 3

MODEL 410VP

CALIBRATION

SETUP AND CALIBRATION

The sensor is calibrated at the factory and does not require initial user-calibration. Simply configure the analyzer to accept a four-electrode sensor and enter the cell constant and calibration factor printed on the label.

After a period of service, the sensor may require calibration. The sensor can be calibrated against a solution having known conductivity or against a referee meter and sensor. If using a standard solution, choose one having conductivity greater than 500 uS/cm. Do not use standard solutions having conductivity less than 100 uS/cm. They are susceptible to contamination by atmospheric carbon dioxide, which can alter the conductivity by a variable amount as great as 1.2 uS/cm (at 25°C). Calibration changes the cell constant only, not the calibration factor. If you wish to change the calibration factor, consult the factory. For more information about calibrating contacting conductivity sensors see application sheet ADS 43-024 available on the Rosemount Analytical website at www.rosemountanalytical.com.

TROUBLESHOOTING

PROBLEM

PROBABLE CAUSE

SOLUTION

 

Wiring is incorrect.

Verify wiring.

 

RTD is open or shorted.

Check RTD for open connections or

 

shorts. See Figure 1.

 

 

Off-scale reading

Sensor is not in process stream.

Be sure sensor is completely submerged

 

 

in process stream.

 

 

 

Variopol cable is not properly seated.

Loosen connector and reseat.

 

Sensor is improperly installed in process

Be sure sensor is completely submerged

Noisy reading

stream.

in process stream.

 

Variopol cable is not properly seated.

Loosen connector and reseat.

 

Bubbles trapped on sensor.

Be sure sensor is installed so that air

 

cannot become trapped against it.

 

 

 

 

 

 

 

Check that temperature correction algo-

 

Wrong temperature correction algorithm.

rithm is appropriate for the sample. See

 

 

the analyzer manual for more information.

 

 

 

 

 

Verify that the correct cell constant and

 

Wrong cell constant. Wrong calibration

calibration factor have been entered in the

Reading seems wrong (lower

analyzer. See the analyzer manual for

factor.

or higher than expected)

more information.

 

 

 

 

 

 

 

Bottom of sensor is too close to pipe

Maintain at least 1.0 in (25 mm) clearance

 

between bottom of sensor and opposite

 

wall.

 

pipe wall.

 

 

 

 

 

 

 

Disconnect red and white RTD wires.

 

Temperature reading in error

Measure resistance across leads, which

 

should be about 1100Ω at room

 

 

 

 

temperature.

 

 

 

 

Electrodes are fouled.

Clean electrodes.

Sluggish response

Sensor is installed in dead area in

Move sensor to a location more repre-

 

 

process piping

sentative of the process liquid.

3

Page 3
Image 3
Emerson PN 51A-410VP instruction sheet Setup And Calibration, Troubleshooting, Problem, Probable Cause, Solution