Emerson UT-P+-STXXXX Calibration Procedure, Summary of Main Menu, Enter Main Menu, First Option

Models: MA N-00 05-00 UT-P+-STXXXX

1 44
Download 44 pages 2.25 Kb
Page 30
Image 30
Table 2 - Table of Responses

Summary of Main Menu

Enter Main Menu:

10 second timer count down

First Option:

Calibrate Sensor, followed by the “YES?” prompt for selection

Second Option:

Set Relay Option, followed by the “YES?” prompt for selection

Third Option:

Review Settings, followed by the “YES?” prompt for selection

Exit:

Return to normal (if no option is selected)

The following table of responses gives a detailed description of toxic detector conditions.

Table 2 - Table of Responses

Conditional Codes

Current O/P

Status LED

Status LED

Alphanumeric display

(mA)

RED

GREEN

 

 

Main Menu

 

 

 

 

Calibrate Sensor

3

N/A

N/A

N/A

Set Relay Options

 

 

 

 

Review Settings

 

 

 

 

Start-up delay

3

Slow flash

 

START DELAY

White sensor lead open

2.5

Fast flash

 

SENSOR FAULT

Black sensor lead open

2.5

Slow flash

 

SENSOR FAULT

Red sensor lead open

2.5

Slow flash

 

SENSOR FAULT

Excess drift (>10%)

2.5

Blip/ blink

 

NEG DRIFT

Auto zero set

3

 

Solid

SETTING ZERO

Apply calibration gas

3.3

Fast flash

 

APPLY 50% of full scale

 

 

 

 

span gas

Span is set, remove gas

3.6

 

Solid

REMOVE GAS

Return to normal operation

3.6

 

Solid

CAL COMPLETE

Normal

4

 

Blip/ blink

0

Gas present

4.40 - 20.00

Blip/ blink

 

0 to 100 of full scale

Calibration Procedure

NOTE

The Toxic Detector should always be calibrated when first installed in the field and prior to

calibration we recommend that the toxic detector run for at least 4 hours.

24

Page 30
Image 30
Emerson UT-P+-STXXXX Calibration Procedure, Summary of Main Menu, Enter Main Menu, second timer count down, First Option