Summary of Main Menu
Enter Main Menu: | 10 second timer count down |
First Option: | Calibrate Sensor, followed by the “YES?” prompt for selection |
Second Option: | Set Relay Option, followed by the “YES?” prompt for selection |
Third Option: | Review Settings, followed by the “YES?” prompt for selection |
Exit: | Return to normal (if no option is selected) |
The following table of responses gives a detailed description of toxic detector conditions.
Table 2 - Table of Responses
Conditional Codes | Current O/P | Status LED | Status LED | Alphanumeric display | |
(mA) | RED | GREEN | |||
|
| ||||
Main Menu |
|
|
|
| |
Calibrate Sensor | 3 | N/A | N/A | N/A | |
Set Relay Options | |||||
|
|
|
| ||
Review Settings |
|
|
|
| |
3 | Slow flash |
| START DELAY | ||
White sensor lead open | 2.5 | Fast flash |
| SENSOR FAULT | |
Black sensor lead open | 2.5 | Slow flash |
| SENSOR FAULT | |
Red sensor lead open | 2.5 | Slow flash |
| SENSOR FAULT | |
Excess drift (>10%) | 2.5 | Blip/ blink |
| NEG DRIFT | |
Auto zero set | 3 |
| Solid | SETTING ZERO | |
Apply calibration gas | 3.3 | Fast flash |
| APPLY 50% of full scale | |
|
|
|
| span gas | |
Span is set, remove gas | 3.6 |
| Solid | REMOVE GAS | |
Return to normal operation | 3.6 |
| Solid | CAL COMPLETE | |
Normal | 4 |
| Blip/ blink | 0 | |
Gas present | 4.40 - 20.00 | Blip/ blink |
| 0 to 100 of full scale |
Calibration Procedure
NOTE
The Toxic Detector should always be calibrated when first installed in the field and prior to
calibration we recommend that the toxic detector run for at least 4 hours.