Calibration Procedure
NOTE
The Toxic Detector should always be calibrated when first installed in the field and prior to calibration we recommend that the toxic detector run for at least 4 hours.
It is necessary to calibrate the
Calibration Procedure for
<Confirm that the system is
<Ensure that the sensor is in a clean air environment before beginning the calibration procedure or we recommended to flow certified ZERO AIR at a rate of 0.5 litres per minute through the barbed tubing connector on the end of the sensor calibration cup accessory, for one minute to ensure clean air is present. If the input device is another transmitter be sure that it is calibrated and that the input to the Toxic Detector is 4.0mA.
<Enter the main menu by activating the magnetic reed switch or manual push button switch. When the countdown from 10 to 0 is complete remove the magnet from the magnetic reed switch or release the CAL/RESET push button. When prompted with Calibrate sensor “YES?” select the function by momentarily placing the magnet on the reed switch or depressing the CAL/RESET button. The selection is acknowledged with flashing “YES”. The controller starts calibration. The following Instructions are scrolled across the display for calibration:
Setting Zero - At this time input should be 4.0mA or ZERO gas applied to the sensor (3.0mA output current level).
Apply 50 % Span - at this time input should be 12.0mA or apply 50% span gas at a rate of 0.5 litres per minute to the end of the sensor calibration cup accessory (3.3mA output current level).
Figure 15 - Magnetic Reed Switch
Activation
Setting Span - increasing input signal has been detected (3.3mA output current level).
Remove Gas - at this time input should be reduced to 4.0mA by removing span gas from the sensor (3.6mA output current level).
Cal Complete - Returns to normal operation (4.0mA output current)
NOTE:
If span setting does not complete successfully within ten minutes of starting the calibration
21