Flow Meter FC01-LQ
5 Configuration (menu option: CONFIGURATION)
The CONFIGURATION menu serves to adjust the
5.1 Selection of monitoring head (menu option: SENSOR SELECT)
This menu option comprises all monitoring heads that can be used with the
The appropriate sensor can be selected from the following menu (for type designation see type label or CPI catalogue).
• TYPE CSP11WM1 | (insertion head for water with sensor adapter |
• TYPE CSP11WM1BV | (insertion head for water with ball valve |
• TYPE CSF11WM1 | (monitoring head for water with adjustable immersion depth) |
•TYPE
Caution!
Menu option SENSOR SELECT may influence data in the parameter selection menu (see para. 5.12, Quitting the configuration menu).
5.2 Monitoring head data (menu option: SENSOR CODE)
To operate the
These characteristics are specified by the sensor code which together with the monitoring head type number is marked on the monitoring head housing.
Enter the following characteristics when selecting a CSF monitoring head. The setting is menu driven:
1. | SENSOR CODE C value | characteristics range: 700 … 1300 |
2. | SENSOR CODE T value | characteristics range: 01 … 99 |
Caution!
Take care to repeat these settings after replacing the monitoring head or electronic control unit
Enter the following characteristics when selecting a CSP monitoring head:
1. | SENSOR CODE C value | characteristics range: 700 … 1300 |
2. | SENSOR CODE value | characteristics range: 01 … 99 |
The nominal diameter of the ball valve is selected in menu option BV SIZE SELECT.
Available sizes are:
• DN25 (1 in), DN32 (1 1/4 in), DN40 (1 1/2 in) und DN50 (2 in).