8.5 Metrological data

8.5.1 Flow rate measurement

Measuring is possible up to the flow rates indicated in the display range. However, the indicated accuracy is no longer guaranteed. The repeatability value remains valid.

8.5.1.1 Monitoring head CSP- with sensor adapter type TP-..

Flow rate measurement:

Sensor adapter

Measuring range

Display range

type

 

in m3/h

in m3/h

TP01

0.02

2.2

2.9

TP02

0.04

3.4

4.5

 

 

 

 

 

TP03

0.05

5.3

7.1

 

 

 

 

 

TP04

0.1

8.7

11.6

 

 

 

 

TP05

0.14

… 13.6

18.1

 

 

 

 

TP06

0.2

… 21.2

28.3

 

 

 

 

 

8.5.1.2 Monitoring head CSF

Flow measurement ranges:

The flow measurement range is determined by the inner pipe diameter (see table). It can be calculated with the following equation:

Q = VN x AR

Q [m3/h] - flow quantity

VN [m/h] - average standard velocity

AR [m2] - inner pipe cross-section

Technical data

53