Operation
Set Up 3
Section 3
Operation
3.1 Set Up
The only
When power is first applied to the unit, the alarm and one or more of the LED's may light. After a maximum of three seconds, the indicators will extinguish and the alarm will reset.
The unit should now be operational with the OPERATION INDICATOR LED flashing on or off with each background pulse received.
The Primalert 35 is a fully digital preset count instrument. The measurement and reset time are dependent on count rate. Refer to the response time specifications listed in Section 2.
3.2 Testing
Test the unit for proper operation by positioning the check source (P/N
3.3 Setting the Alarm Level
The Alarm Level may be set with a small screwdriver from the front panel. Alarm levels from 1 to 32 mR/h may be set. When the radiation exceeds the setting, the alarm will be actuated.
3.4 Alarm Reset
The alarm reset on the Primalert 35 is automatic and both aural and visual alarms will stop when the radiation level drops below the Alarm Level setting.