Documenting Process Calibrator HART Mode

Cloning a Transmitter

Cloning a Transmitter

Cloning copies the Basic Setup information from one transmitter to another. You can clone generic and supported transmitters.

To use the calibrator to clone a transmitter:

1.Connect the calibrator to the transmitter you want to clone. You only need a communication connection.

2.From the Active Device screen, press the Setup softkey.

3.Use the uor dkeys to select Basic from the list on the screen and press e.

4.Press the Store Page softkey to load the settings into the calibrator’s memory (not the transmitter).

5.Disconnect the calibrator from the transmitter and connect it to the transmitter you want to configure. Press the Abort softkey to establish communication.

6.Press the Setup softkey, then select Basic to return to the Basic Setup screen.

7.Press the Recall Page softkey to refresh the calibrator display with the cloned parameter settings. The settings are not transmitted to the transmitter at this time.

8.Use the uor dand ekeys to select parameters one at a time, change them, or leave them as cloned, and press the Send softkey to transmit them to the transmitter.

pe09s.bmp

19

Page 27
Image 27
Fluke 744 manual Cloning a Transmitter