12Multimeter Users Manual

Use i only on power supplies and other power sources that have a low output impedance. Do not use i to measure voltage in electronic circuitry unless a 2 ke load will not damage the circuit. See † on page 27.

DISABLING iWITH FUNCTION LOCK

To lock the meter in either the continuity/diode or ohms function, and disable i:

1.Put the slide-switch in the continuity/ohms position. The meter selects the continuity/diode or ohms function. Press [g] to toggle between the continuity/diode and ohms functions.

2.Press [Z] to put the meter in manual range. Z is displayed. The meter is locked in the selected function and i is disabled.

In continuity/diode test, press [Z] to toggle the beeper on and off. In ohms, press [Z] to manually select a range.

To remove the function lock and reenable i, press [Z] for 2 seconds, press [g], or move the slide-switch.

14

Page 17
Image 17
Fluke PN 2063508 user manual Disabling iWITH Function Lock