Filament Voltage Test.

Filament Voltage Test.

When replacing either the flyback transformer or the video board, the filament voltage may not be correct.

Measuring the filament voltage is not accurate using a true RMS voltage meter, because of the high frequency components, which make up the filament voltage. An oscilloscope, with RMS capability, may be used to measure the filament voltage.

Although a visual check of the filament color is a indicator of the filament voltage, it is a good practice to check the filament voltage if there is any doubt about this important monitor parameter.

The following filament voltage test is an accurate method of finding the true RMS voltage to the filament. This is accomplished by comparing the light output of the filament when it is driven by the monitor to the light output of the filament with an applied DC voltage using a loaded photocell.

SETUP:

 

 

 

903

 

62K,912

 

 

 

 

 

 

 

 

 

 

 

ISOLATED

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DUAL 1A DC

 

10

 

 

 

0

, 904

 

 

 

 

 

 

 

 

 

0, 902

 

 

1N4005

 

 

 

 

 

 

POWER

 

 

11

 

 

 

 

 

 

+

 

 

 

 

250V

1N4005

 

 

 

 

906

10uF+

 

 

SUPPLY

 

 

 

 

 

1N4005

 

 

 

 

 

 

.015uF

 

 

 

18

, 879

895

10uF

0 to 30V

0 to 30V

+

+

 

13

 

 

905

 

 

 

 

 

8

 

FIL.

FIL.

10

2.2nF,889

 

 

 

 

 

 

14

 

901

 

 

 

 

 

0

, 896

11

 

 

 

 

 

 

RED

 

 

CF

 

RED

BLUE

 

 

 

 

 

205

15

 

 

 

 

0

, 897

 

 

 

 

 

 

 

 

,

 

 

 

 

 

12

 

 

 

 

 

16

 

7

 

0

, 898

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+6.30V

 

 

 

 

17

 

 

470

 

G2

 

 

 

9

10

 

 

GAPS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

18

 

900

 

 

 

 

 

8

11

 

GND

 

DVM

 

 

 

 

H400

 

 

 

 

 

 

7

 

12

 

 

 

 

 

19

 

 

6

 

 

 

 

 

 

 

+

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

20

 

 

 

 

 

6

CELL

 

 

 

 

 

 

 

 

 

899

GREEN

 

 

 

 

 

 

 

 

.150V

 

 

 

 

 

 

 

 

 

5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SOCKET

 

 

 

DVM

 

470, CF

 

 

 

 

 

 

 

 

 

 

 

 

 

5

 

 

 

 

 

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

62K,852

 

 

 

 

G1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FR205

 

 

 

 

 

 

 

877

CRT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

853

 

 

1K

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

XX93

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

.033uF, 854

CF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Video Board

 

 

 

 

100K, CF

855

 

 

 

 

 

 

 

 

 

24.3K

 

 

 

 

 

 

 

 

 

 

 

 

 

To find the filament voltage;

1.Solder two short buss wires to the filament pins to clip on.

2.Use black tape to secure the photocell over the hole in the plastic CRT socket connector. Caution: Be sure not to move the photocell between the two tests.

3.Record the DC voltage output from the photocell with the monitor running normally. The monitor should be powered up for 10 minutes before making this measurement.

4.Turn off the monitor.

5.Connect the variable voltage, 1 amp, DC power supply to the filament. The negative lead to the filament ground at the CRT socket pin 10.

6.Adjust the power supply voltage for the photocell reading, recorded in step 3.

7.The equivalent filament RMS voltage is now recorded by measuring the DC voltage at the CRT socket pins 9 and 10.