ELECTRICAL CONNECTIONS
Line voltage supplied to the cabinet junction box must not be affected by the operation of other electrical equipment. Junction box is located at the rear of the lower section. The rear compressor cover is maintained in place with Velcro. Pull to remove it.
WARNING: ELECTRICAL AND GROUNDING CONNECTIONS MUST COMPLY WITH THE APPLICABLE PORTIONS OF THE NATIONAL ELECTRICAL CODE AND/OR OTHER LOCAL ELECTRICAL CODES.
WARNING: DISCONNECT ELECTRICAL POWER SUPPLY AND PLACE A TAG AT THE DISCONNECT SWITCH INDICATING THAT YOU ARE WORKING ON THE CIRCUIT.
The circuit should be protected with a Dual Element
ELECTRICAL DATA
|
| Minimum Circuit Ampacity | |
Model | Volts / Hertz / Phase | Maximum Protective Device | |
|
| AMPS | |
|
|
| |
HQC45 | 208 / 60 / 1 | 15 | |
|
| ||
HQCF45 | 240 / 60 / 1 | 15 | |
| |||
|
|
| |
| 208 / 60 / 1 | 20 | |
|
|
| |
HQC90 | 240 / 60 / 1 | 20 | |
|
| ||
208 / 60 / 3 | 15 | ||
| |||
|
|
| |
| 240 / 60 / 3 | 15 | |
|
|
|
Compiled in accordance with the National Electrical Code, NFPA 70, latest edition.
INSTALLATION OF REMOTE ALARM (Optional)
The chiller provides a connection for a remote alarm that operates when the buzzer sounds on completion of a chill cycle. Connect to terminals on terminal block (# 5 and # 6 in the upper section of the chiller) per the wiring diagram and these restrictions:
1.Maximum remote alarm rating: 120V - 240V at 2 amps resistive or 100 watt incandescent lamp.
2.Connection must be made of 600 volt insulated wire suitable for supply voltage. Do not use bell wire, lamp cord or similar type wire.
INSTALLATION CHECKOUT
The
After proper electrical connections are made, place the switch next to the control in the On position.