HoMedics BPA-300 Potential for RADIO/TELEVISION Interference, Troubleshooting, For USA only

Models: BPA-300

1 34
Download 34 pages 20.57 Kb
Page 16
Image 16

BPA-300-A.qxd 10/5/06 4:20 PM Page 29

POTENTIAL FOR RADIO/TELEVISION

INTERFERENCE

For USA only:

This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. The product generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If the product does cause harmful interference to radio or television reception, which can be determined by turning the product on and off, the user is encouraged to try to correct the interference by one or more of the following measures:

Reorient or relocate the receiving antenna.

Increase the separation between the product and the receiver.

Connect the product into an outlet on a circuit different from that to which the receiver is connected.

Consult the dealer or an experienced radio/TV technician for help.

For Canada only:

This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus as set out in the interference-causing equipment standard entitled “Digital Apparatus”, ICES-003 of the Canadian Department of Communications.

Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux appareils numériques de Classe B prescrites dans la norme sur le matériel brouilleur: “Appareils Numériques”, ICES-003 édictée par le ministre des communications.

TROUBLESHOOTING

If any abnormality arises during use, please check the following points

SYMPTOMS

POSSIBLE

CORRECTION

CAUSES

 

 

 

 

 

Unit does not turn on

Batteries have

Replace them with

when the

run down.

four new AA alkaline

START/STOP

 

batteries.

button is pushed.

 

 

Battery polarities

Re-insert the batteries

 

have been positioned

in the correct

 

incorrectly.

positions.

 

 

 

EE measurement

Cuff has been placed

Wrap the cuff

error symbol shown

incorrectly.

properly so that it is

on display or the

 

positioned correctly.

blood pressure value

 

 

Did you talk or move

Measure again.

is displayed

during measurement?

Keep arm steady

excessively low

 

during measurement.

Shaking of the arm

(or high).

with the cuff on.

 

 

 

 

 

 

E1 error symbol

Air circuit abnormality.

Check cuff connection.

shown on display

Cuff tube may not be

Measure again.

 

plugged into monitor

 

 

correctly.

 

 

 

 

E2 error symbol

Inflation pressure

Turn the unit off, then

shown on display

exceeding 300 mmHg.

measure again.

 

 

 

E3 error symbol

Data error.

Remove and reload

shown on display

 

the batteries.

 

 

 

CA appears on

Recalibration alert.

Refer to page 26 for

display

more information.

 

 

 

 

Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page.

29

30

Page 16
Image 16
HoMedics BPA-300 warranty Potential for RADIO/TELEVISION Interference, Troubleshooting, For USA only, For Canada only