BATTERY IN USE |
|
|
| ||
P = Primary (“D” cell) |
|
| CURRENT GAS CONCENTRATION | ||
S = Standby (Internal NiCad) |
| ||||
| (Blinking indicates sensor over range) | ||||
Flashing indicates low battery |
| ||||
|
| ||||
|
|
|
|
| TARGET GAS |
|
|
| 0 |
| |
| P | PPM |
| ||
|
|
|
| NH3 |
|
| A |
|
|
| |
|
|
|
|
| |
OFF | MENU SMPL | Battery Light |
Power
ADVANCED FEATURES ONLY
ALARM INDICATION CODEDo not push for normal operation
Blank = No alarm condition present
C =
W =
A =
P = Pump trouble
Lg = Data logging in progress
Figure 2. Normal Operating Screen
2.3 Response Test
Prior to use, the unit should first be tested for proper response. With the detector operating, the unit should be exposed to a sample of the target gas.The display should show an increasing concentration. If not, do not use it for field measurements.
2.4 Sample Inlet Port
The Honeywell
2.5 Response/Recovery Time
Under normal conditions, the instrument will reach 90% of final value within two minutes. This is dependent on concentration and temperature.
Recovery time for the sensor depends on duration and concentration of exposure. Short exposures of concentrations at the lower end of the target gas range result in rapid recovery. Long exposures to levels above the middle of the target gas range or short exposures to levels exceeding the target gas range can extend recovery times to hours. Repeated exposures above the target gas range will reduce cell life and should be avoided.
10 | Honeywell |