Warning Function
4.Push WARN button.
5.Push FUNC (function) button.
6.Push SELECT button while viewing the arrow in the lower left hand corner of the screen to toggle through the following three possibilities:
•UP arrow — Warn on increasing concentration (gas goes higher than setpoint).
•DOWN arrow — Warn on decreasing concentration (gas goes lower than setpoint).
•Blank — warning function turned OFF.
7.Push SAVE button when the DOWN arrow is displayed to store new warning function.
8.Push DONE button to return to Alarms mode.
To Exit CAUT / WARN / ALRM Mode
9.Push DONE button two times slowly to get back into normal operation mode.
The Caution and Alarm functions are programmed in the same manner, except for step four above in which you would push either the CAUT or ALRM button.
4.2 Display Variables
The Honeywell
To Enter AVG / BLANK Mode
1.Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).
2.Push and hold the unnamed button first, then Push and hold the Menu button. Hold both buttons until the screen goes blank (two seconds). Release both buttons and the unit is now in Programming Mode.
3.Push the MORE button.
4.Push the DISP (display) button and the unit is now in the AVG / BLANK mode.
16 | Honeywell |