Honeywell EC-P2 Vent through flowmeter Bypass Tee, Calibration gas flowmeter, Lpm Air

Models: EC-P2

1 28
Download 28 pages 24.6 Kb
Page 14
Image 14
LPM AIR

External flowmeter indicates no flow: Verify that pump is running. Check that the manifold screws are tight and the extension wand connection is tight.

Unit does not respond when keys are pressed: Press and hold all four buttons for five seconds. This will reset the Honeywell EC-P2. Turn the unit back on by pressing the PWR button.

High reading won’t clear: When the sensor is exposed to very high gas concentrations, it takes a relatively long time for the sensor to clear. Leave the unit running in a clean environment for a few hours to clear the sensor. If it doesn’t clear after six hours, the sensor may have been damaged. Contact Honeywell Analytics for a replacement SMART-CELL.

Unit is in alarm with gas reading of 0 ppm: Check the alarm setpoints and alarm functions as described on page 12. Verify that alarm is not programmed as a downscale alarm, and that the setpoint is above zero.

1.0

.8

.6

.4

.2

LPM AIR

Vent through flowmeter

Bypass Tee

Calibration gas flowmeter

TEST UNIT PRIOR TO USE

OUTLETVent through flowmeter

INLET

CAUTION

DO NOT BLOCK INLET OR OUTLIET PORTS

Calibration gas bottle

(use only Certified Calibration Gas)

Figure 4. Calibration of the Honeywell EC-P2 Sensor using a Bypass Tee

14

Honeywell EC-P2 Instruction and Installation Manual

Page 14
Image 14
Honeywell EC-P2 Vent through flowmeter Bypass Tee, Calibration gas flowmeter, Calibration gas bottle, Lpm Air