To calibrate a detector, complete the following:

MicroDock II

Quick Reference Guide

A calibration adjusts the sensor sensitivity to ensure an accu- rate response to gas.

To calibrate a detector, complete the following:

1.From the docking module of the applicable detector, press C CALIBRATION.

2.The gas is automatically applied.

3.When the calibration is complete, the system purges. The countdown for time remaining displays beside Purge.

Manual background = calibration has passed

Manual background = calibration has failed

If any test fails, refer to Troubleshooting in the

MicroDock II Base Station User Manual.

4.The detector deactivates.

Data Transfer (GasAlert Extreme, GasAlertMicroClip, GasAlertMax XT only)

This feature transfers datalogs from the detector to the base station MultiMediaCard (MMC).

To transfer a datalog, complete the following:

1.Activate the detector and insert it into the docking module.

2.Press C DATA TRANSFER. The RUN LED flashes rapidly. The base station LCD displays the following screen:

Depending upon how many logs are stored on the MMC, the data transfer may require 12-14 minutes to complete (1-2 minutes for the GasAlertMicroClip and GasAlertMax XT).

Note

A maximum of ten datalogs from all modules combined can be stored on an MMC (100 datalogs for GasAlertMicroClip and GasAlertMax XT).