Honeywell MicroDock II manual Charging the Battery Pack Optional, Quick Reference Guide

Models: MicroDock II

1 19
Download 19 pages 40.66 Kb
Page 14
Image 14
Charging the Battery Pack (Optional)

MicroDock II

Quick Reference Guide

GasAlertMicroClip

GasAlertMax XT

The detector can be configured prior to performing a bump test or calibration to

change user options,

sensor options, and

program options.

The detector can also be configured to disallow bump tests or calibrations unless performed with the MicroDock II. For more information, refer to Configuring Detectors in the Fleet Manager II Technical Reference Guide.

Charging the Battery Pack (Optional)

aWarning

There is a maximum of six charging docking modules can be installed on the MicroDock II base station (six charging modules plus four non-charging modules for a maximum of ten docking modules per station).

To charge successfully, the temperature must be between 50°F to 95°F (10°C to 35°C). Charge the battery immediately when the detector emits a low battery alarm.

aCaution The charger is for indoor use only.

Note

The detector will not charge while bump tests or calibrations are being performed. If the battery is low,

charge the detector for 30 minutes, then begin a test. When the test is complete resume charging the battery.

When charging a new battery for the first time, ensure the bat- tery obtains a full charge. For more information, refer to the detector’s user manual or operator’s manual.

Note

Up to six GasAlertMicro and GasAlertMicro 5/PID/IR detectors can charge simultaneously.

Up to four GasAlertMicroClip and GasAlertMax XT can charge simultaneously.

For more information, refer to the MicroDock II Base Station User Manual.

To charge the battery pack, complete the following:

1.Connect the charger adaptor cord into the CHARGE port on the station, and then plug into an AC outlet. The CHARGE LED briefly lights red then green dur- ing the self-test. The LED then powers off (not appli- cable to the GasAlertMicroClip and GasAlertMax XT modules).

2.Deactivate the detector. Insert the detector into the charging cradle.

3.The CHARGE LED lights red (all charging modules).

12

Page 14
Image 14
Honeywell MicroDock II manual Charging the Battery Pack Optional, Quick Reference Guide