A
SpecificationsSpecificationsTable
Parameter |
|
| HP 6631B |
| HP 6632B |
| HP 6633B |
| HP 6634B |
| |||||
|
|
|
|
|
| HP 66332A |
|
|
|
|
|
| |||
Output Ratings | Voltage: | 0 − 8 V |
|
| 0 − 20 V |
| 0 − 50 V |
| 0 − 100 V |
| |||||
|
| Current: | 0 − 10A |
| 0 − 5A |
|
| 0 − 2 A |
|
| 0 − 1A |
|
| ||
Programming Accuracy | Voltage: | 0.05% + 5 mV | 0.05% + 10 mV | 0.05% + 20 mV | 0.05% + 50 mV | ||||||||||
± ° | C) | +Current: | 0.05% + 4 mA | 0.05% + 2 mA | 0.05% + 1 mA | 0.05% + 0.5 mA | |||||||||
(@ 25C 5 |
|
|
|
|
|
|
|
|
|
|
|
|
| ||
DC Measurement | Voltage: | 0.03% + 2 mV | 0.03% + 3 mV | 0.03% + 6 mV | 0.03% + 12 mV | ||||||||||
Accuracy |
| Low Current range |
|
|
|
|
|
|
|
|
|
|
|
| |
(via | − | 0.1% + 2.5 | μ | 1 | 0.1% + 2.5 | μ | 1 | 0.1% + 2.5 | μ | 1 | 0.1% + 2.5 | μ | 1 | ||
meters with respect to | 20 mA to +20 mA: |
| A |
| A |
| A |
| A | ||||||
High Current range |
|
|
|
|
|
|
|
|
|
|
|
| |||
actual output at 25C |
|
|
|
|
|
|
|
|
|
|
|
| |||
+20 mA to +rated I: | 0.2% + 1 mA2 | 0.2% + 0.5 mA2 | 0.2% + 0.25 mA2 | 0.2% + 0.25 mA2 | |||||||||||
±5°C) |
| −20 mA to −rated I: | 0.2% + 1.6 mA | 0.2% + 1.1 mA | 0.2% + 0.85 mA | 0.2% + 0.85 mA | |||||||||
|
| ||||||||||||||
Ripple and Noise | Normal mode voltage | 0.3 mV/3 mV3 | 0.3 mV/3 mV3 | 0.5 mV/3 mV3 | 0.5 mV/3 mV3 | ||||||||||
(in the range of 20 Hz to | |||||||||||||||
20 MHz with outputs | Fast mode voltage |
|
|
|
|
|
|
|
|
|
|
|
| ||
ungrounded or with | 1 mV/10 mV | 1 mV/10 mV | 1 mV/15 mV | 2 mV/25 mV | |||||||||||
either terminal grounded) | |||||||||||||||
Current (rms): | 3 mA |
|
| 2 mA |
|
| 2 mA |
|
| 2 mA |
|
| |||
|
|
|
|
|
|
|
|
|
| ||||||
Load Regulation4 | Voltage: | 2 mV |
|
| 2 mV |
|
| 4 mV |
|
| 5 mV |
|
| ||
(change in output voltage | Current: | 2 mA |
|
| 1 mA |
|
| 1 mA |
|
| 1 mA |
|
| ||
or current for any load |
|
|
|
|
|
|
|
|
|
|
|
|
| ||
change within ratings) |
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Line Regulation | Voltage: | 0.5 mV |
|
| 0.5 mV |
|
| 1 mV |
|
| 1 mV |
|
| ||
(change in output voltage | Current: | 1 mA |
|
| 0.5 mA |
|
| 0.25 mA |
| 0.25 mA |
| ||||
or current for any line |
|
|
|
|
|
|
|
|
|
|
|
|
| ||
change within ratings) |
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Transient Response | Normal mode: | < 100 μs |
|
|
|
|
|
|
|
|
|
|
| ||
Time4 |
| Fast mode: | < 50 μs |
|
|
|
|
|
|
|
|
|
|
|
(for the output voltage to recover to its previous level within 0.1% of the voltage rating of the unit or 20 mV following a change in load current of up to 50% of the output current rating)
1This specification may degrade slightly when the unit is subjected to an RF field ≥3 V/meter.
2Applies in SCPI mode, with current detector set to DC. With current detector set to ACDC, accuracy is 0.2% + four times the fixed error value. In COMPatibility mode, accuracy is 0.2% + six times the fixed error value.
3For HP 6631B, 6632B, and 66332A (from 1 MHz to 20 MHz ) = 0.3mV/15mV; For HP 6633B and 6634B (from 1 MHz to 20 MHz ) = 0.5mV/15mV
4Applies at rear terminals with unit set to remote sensing and with sense terminals externally jumpered to their respective output terminals.