Verification and Calibration - B
Performing the Calibration ProcedureNOTE: The calibration procedure can only be performed using the SCPI language commands. Use either the front panel Address key to access the LANG command, or use the SYSTem:LANGuage command to change the programming language to SCPI.
Table
You do not have to do a complete calibration each time. If appropriate, you may calibrate only the voltage or current and proceed to "Saving the Calibration Constants". However, the voltage or current calibration sequence must be performed in its entirety. The following parameters may be calibrated:
♦voltage programming and measurement
♦overvoltage protection (OVP)
♦current programming and measurement
♦negative current programming
♦low range measurement
♦ac current measurement
Front Panel Calibration Menu
The Entry keypad is used for calibration functions.
Shift Cal
Press this key to access the calibration menu.
Display | Command Function |
CAL ON <value> | Turns calibration mode on when the correct password |
| value is entered. |
CAL OFF | Turns calibration mode off |
CAL:LEV <char> | Advance to next step in sequence (P1 or P2). |
CAL:DATA <value> | Enter an external calibration measurement. |
CAL:VOLT | Begin voltage calibration sequence |
CAL:VOLT:PROT | Begin voltage protection calibration |
CAL:CURR | Begin high range current calibration sequence |
CAL:CURR:NEG | Begin negative current calibration sequence |
CAL:CURR:MEAS:LOW | Begin low range current measurement calibration |
CAL:CURR:MEAS:AC | Begin ac current calibration sequence |
CAL:SAVE | Saves the calibration constants in |
CAL:PASS <value> | Set new calibration password. |
Notes:
value = a numeric value
char = a character string parameter
Use p | and | q | to scroll through the menu commands. |
Use ã | and | ä | to scroll through the menu parameters. |
Use á | and | â | to select a digit in a numeric entry field. |