Verification and Calibration - B

Performing the Calibration Procedure

NOTE: The calibration procedure can only be performed using the SCPI language commands. Use either the front panel Address key to access the LANG command, or use the SYSTem:LANGuage command to change the programming language to SCPI.

Table B-1 lists the equipment required for calibration. Figure B-1 shows the test setup.

You do not have to do a complete calibration each time. If appropriate, you may calibrate only the voltage or current and proceed to "Saving the Calibration Constants". However, the voltage or current calibration sequence must be performed in its entirety. The following parameters may be calibrated:

voltage programming and measurement

overvoltage protection (OVP)

current programming and measurement

negative current programming

low range measurement

ac current measurement

Front Panel Calibration Menu

The Entry keypad is used for calibration functions.

Shift Cal

Press this key to access the calibration menu.

Display

Command Function

CAL ON <value>

Turns calibration mode on when the correct password

 

value is entered.

CAL OFF

Turns calibration mode off

CAL:LEV <char>

Advance to next step in sequence (P1 or P2).

CAL:DATA <value>

Enter an external calibration measurement.

CAL:VOLT

Begin voltage calibration sequence

CAL:VOLT:PROT

Begin voltage protection calibration

CAL:CURR

Begin high range current calibration sequence

CAL:CURR:NEG

Begin negative current calibration sequence

CAL:CURR:MEAS:LOW

Begin low range current measurement calibration

CAL:CURR:MEAS:AC

Begin ac current calibration sequence

CAL:SAVE

Saves the calibration constants in non-volatile memory.

CAL:PASS <value>

Set new calibration password.

Notes:

value = a numeric value

char = a character string parameter

Use p

and

q

to scroll through the menu commands.

Use ã

and

ä

to scroll through the menu parameters.

Use á

and

â

to select a digit in a numeric entry field.

59