Calibration

Introduction

The power supply may be calibrated either from the front panel or from a controller over the GPIB. The procedures given here apply to all models.

Important These instructions do not include verification procedures. If you need to perform verification as a prerequisite to or as part of your calibration procedure, see “Appendix B - Verification”.

Equipment Required

The equipment listed in Table B-1, or equivalent, is required for calibration.

Table B-1. Equipment Required For Calibration

Equipment

Voltmeter

Shunt resistor

GPIB Controller

Characteristics

Dc accuracy 0.005%, 6 digits

100 A, 0.001 Ω, 0.04%, 100 W

For Calibration over the GPIB

Recommended Model

Agilent 3456A or 3458A

Guildline 9230/100

HP Vectra (or IBM compatible) with GPIB Interface, or Agilent BASIC series

General Procedure

Because the power supply output must be enabled during calibration, voltages or currents hazardous to personnel and/or damaging to equipment can appear at the output terminals.

Parameters Calibrated

The following parameters may be calibrated:

Output voltage.

Output voltage readback.

Overvoltage protection (OVP).

Output current.

Output current readback.

You do not have to do a complete calibration each time. If appropriate, you may calibrate only the voltage or current and proceed to "Saving the Calibration Constants".

Test Setup

Figure B-1 shows the test setups required for voltage and current calibration for each power supply series.

Calibration 97