Digital Subsystem

This subsystem programs the control port on the back of the power supply when it is configured for Digital I/O operation.

DIG:DATA

This command sets and reads the power supply digital control port when that port is configured for Digital I/O operation. Configuring of the port is done via an internal jumper (see Appendix F). The port has three signal pins and a digital ground pin. Pins 1 and 2 are output pins controlled by bits 0 and 1. Pin 3 is controlled by bit 3 and can be programmed to serve either as an input or an output. Pin 4 is the digital ground.

Bit position 2 normally serves as an output. To change it to an input, it must first be programmed high. The DIG:DATA? query returns the last programmed value in bits 0 and 1 and the value read at pin 3 in bit 2. The bits are turned on and off in straight binary code as follows:

Value 0 1 2 3

Bit

Configuration

0

1

2

0

0

0

1

0

0

0

1

0

1

1

0

 

 

 

 

Digital I/O Port Programming Chart

 

 

Pin Configuration1

 

 

Bit

 

 

 

Configuration

1

 

2

3

4

Value

0

1

2

Lo

 

Lo

 

Output

Gnd

4

0

0

1

Hi

 

Lo

 

Output

Gnd

5

1

0

1

Lo

 

Hi

 

Output

Gnd

6

0

1

1

Hi

 

Hi

 

Output

Gnd

7

1

1

1

 

 

 

 

1Pins 1

and 2 are

always outputs

 

 

Pin Configuration1

1

2

3

4

Lo

Lo

Input

Gnd

Hi

Lo

Input

Gnd

Lo

Hi

Input

Gnd

Hi

Hi

Input

Gn

Command Syntax

[SOURce]:DlGital:DATA[:VALuel <NRf>

Parameters

0 to 7

Suffix

(None)

*RST Value

0

Examples

DIG:DATA 7 DIGITAL:DATA:VALUE 7

Query Syntax

[SOURce]:DlGital:DATA?

Returned Parameters

<NRI> Values from 0 to 7

Related Commands

*RST *RCL *SAV

Display Subsystem

This subsystem controls the state and output of the alphanumeric portion of the display.

DISP

Enables or disables the display. When disabled, the display characters are blank. The annunciators are not affected by this command.

Command Syntax

DISPlay[:WINDow][:STATe] <bool>

Parameters

0 1 OFF ON

Suffix

(None)

*RST Value

ON

Examples

DISP ON DISPLAY:STATE ON

Query Syntax

DISPlay[:WINDow][STATe]?

Returned Parameters

<NR1> 0 or 1

Related Commands

DISP:MODE DISP:TEXT *RST

74 Language Dictionary