The power cord supplied with power supply may or may not include a power plug (see "Options" in Chapter l) at one end of the cord. Terminating connections and a ground lug are attached to the other end of the cord. See Figure 2-2 and proceed as follows:

1.If they are not already in place, position the strain relief connector ), safety cover , rubber boot , and

connector nuton the power cord .

2.Secure the ground wire to the chassis earth ground stud.

3.For single-phase operation, connect the neutral wire to the N input terminal and the line wire to the L input terminal (this line is fused inside the unit).

4.For line-to-line operation from a three-phase source as shown in Figure 2-3, connect one phase to the N input terminal and another phase to the L input terminal (this line is fused inside the unit).

Note The N terminal is not internally grounded.

5.Position the safety cover over the power input terminals and tighten the cover screws and strain relief connector

screws .

Figure 2-2. Connecting Power Cord

Figure 2-3. Connection to a 3-Phase Line

Installation 19