Recovering From Calibration Problems

You can encounter serious calibration problems if you cannot determine a calibration password that has been changed or the power supply is severely out of calibration. There are jumpers inside the power supply that permit the calibration password to be defeated and allow the original factory calibration constants to be restored. These jumpers are explained in the Service Manual.

Calibration Error Messages

Error messages that can occur during calibration are shown in Table A-3.

Table B-3. GPIB Calibration Error Messages

Error

No.

1

2

3

4

5

Meaning

CAL jumper prevents calibration1 CAL password is incorrect CAL mode is not enabled

Incorrect computed readback constants Incorrect computed programming constants

Error

Meaning

No.

 

6Wrong CAL command sequence

7Incorrect state (CV/CC) for this command

1This is a hardware disable. See the power supply Service Manual.

Calibration Over The GPIB

You can calibrate the power supply by using SCPI commands within your controller programming statements. Be sure you are familiar with calibration from the front panel before you calibrate from a controller. The SCPI calibration commands are related to the front panel calibration controls as follows:

Front Panel

Command

Corresponding SCPI

Command

CAL:STAT {ON1},<password>

CAL:STAT {OFF0}

CAL:PASS <NRf>

CAL:VOLT:LEV {MINMAX} CAL:VOLT[:DATA] <Nrf>

Front Panel

Command

Corresponding SCPI

Command

CAL:CURR:LEV {MINMAX} CAL:CURR[:DATA] <NRf>

CAL:CURR:MON<newline> CAL:CURR:DATA <NRf>

CAL:SAVE

CAL:VOLT:PROT

Calibration Example

A sample calibration program is given at the end of this appendix. If your system is Agilent BASIC, you can use the program with very little modification. Otherwise, use it as a guide for writing your own program.

100 Calibration

Page 100
Image 100
Agilent Technologies E4356A manual Calibration Over The Gpib, Calibration Example, Error Meaning