Specifications

Specifications are performance parameters warranted over the specified temperature range.

Supplemental Characteristics are not warranted but are descriptions of performance determined either by design or type testing.

Table A-1. Performance Specifications for Agilent E4356A

Parameter

Output Ratings (@ 0 to 45°C)

Programming Accuracy (@ 25 °C ± 5 °C)

Ripple & Noise

(from 20 Hz to 20 MHz with outputs ungrounded, or with either output terminal grounded)

Voltage:

Current:

Voltage:

Current:

Constant Voltage rms: Constant Voltage p-p: Constant Current rms:

(measured with 60 cm leads )

Value

0 - 80 V (0 to 26A)

0 - 30 A (0 to 70 V)

0.04% + 80 mV

0.l % + 25 mA

2mV 16 mV 25 mA

Readback Accuracy

(from front panel or over GPIB with respect to actual output @ 25 °C ± 5 °C)

Voltage:

0.05% + 120 mV

Current:

0.l % + 35 mA

Load Regulation

(change in output voltage or current for any load change within ratings)

Voltage: 0.002% + 3 mV

Current: 0.005% + 2 mA

Line Regulation

(change in output voltage or current for any line change within ratings)

Voltage: 0.002% + 3 mV

Current: 0.005% + 2 mA

Transient Response Time

< 900 μs

(for the output voltage to recover to its previous level within 0.1% of the rated voltage or 20 mV, whichever is greater, following any step change in load current up to 50% of the rated current)

Specifications 93

Page 93
Image 93
Agilent Technologies E4356A manual Specifications, Value