G
Compatibility Language
Introduction
This power supply is programatically compatible with the Agilent 603xA Series AutoRanging Power Supplies (ARPS). This means that you can program this power supply over the GPIB using the ARPS commands. Software that you have written for the autoranging power supplies can also be adapted to program this power supply.
Note | The Agilent E4356A Power supply’s serial link is not supported by ARPS commands. You can use only a |
| GPIB primary address for the power supply. |
|
|
To switch from SCPI commands to ARPS commands (and vice versa), use the SYST:LANG command. This command is documented in Chapter 7.
Table
Parallel Polling
When programmed for parallel polling and Compatibilty Language, power supplies operating under the Agilent BASIC system can "hang up" the GPIB when the system is turned on. This can occur under the following conditions:
∙The controller uses CS80 Protocol for an external disk drive (for example, the Agilent 9133D).
∙The external disk drive and the power supply have the same select code and that code is 7 or less.
∙The external disk drive and power supply addresses are binary complements of each other (e.g., 0 & 7, 1 & 6, etc.)
When the system is turned on, the power supply accesses the GPIB before the controller and prevents it from accessing the external disk drive. The solution is to change one of the GPIB addresses, or to ensure that the power supply is not turned on until after the controller has completed its selftest and has control of the GPIB.