Query Syntax

*STB?

 

 

 

Returned Parameters

<NR1>

(Register binary value)

 

Related Commands

(None)

 

*TRG

Meaning and Type

Trigger Device Trigger

Description

This command generates a trigger when the trigger subsystem has BUS selected as its source. The command has the same affect as the Group Execute Trigger (<GET>) command.

Command Syntax

*TRG

Parameters

(None)

Query Syntax

(None)

Related Commands

ABOR CURR:TRIG INIT TRIG[:IMM] VOLT:TRIG

 

<GET>

*TST?

Meaning and Type

Test Device Test

Description

This query causes the power supply to do a self-test and report any errors (see "Selftest Error Messages" in "Chapter 3 - Turn-On Checkout").

Query Syntax

*TST?

 

Returned Parameters

<NR1>

 

 

0

Indicates power supply passed self-test.

 

Nonzero

Indicates an error code.

Related Commands

(None)

 

*WAI

Meaning and Type

Wait to Continue Device Status

Description

This command instructs the power supply not to process any further commands until all pending operations are completed. "Pending operations" are as defined under the *OPC command. *WAI can be aborted only by sending the power supply a GPIB DCL (Device Clear) command.

Command Syntax

*WAI

Parameters

(None)

Query Syntax

(None)

Related Commands

*OPC *OPC?

Language Dictionary 71

Page 71
Image 71
Agilent Technologies E4356A manual Trg, Tst?, Wai