Current Programming and Readback Accuracy
This test verifies that the current programming and readback are within specification. Connect the appropriate current monitoring resistor (see Table
Table
1
2
3
4
5
6
7
Action
Turn off the power supply and connect the current monitoring resistor as shown in Figure
sufficient size to carry the maximum rated current of the unit (see Table
Connect a DVM across the resistor at the monitoring terminals.
Turn on the power supply and program the output for 5V, 0A.
Disable the output ( )
Enable the output or OUTP ON).
Observe the DVM voltage reading. Divide this by the resistance of the current monitor resistor. Record the result as the Low Current value in applicable test table.
Record the front panel display readback
Normal Result
Value within Low Current limits (see applicable test table).
Value within specified readback limits (see applicable test table).
8
9
10
11
Program output current to full scale.
Repeat Steps 6 and 7.
Disable the output ( ).
Remove the short from across the load.
Both current readings within specified High Current and readback limits (see applicable test table).
Table
Test Description
Minimum Spec
Results *
Maximum Spec
Measurement Uncertainty
Low Voltage (0 V) Vout
Front Panel Display Readback
High Voltage ( 80 V) Vout
Front Panel Display Readback
Low Current (0 A) Iout
Front Panel Display Readback
High Current (30 A) Iout
Front Panel Display Readback
Voltage Programming and Readback
____________ mV | ||
Vout | ____________ mV | |
79.888 V | _____________V | |
Vout | ____________ mV | |
Current Programming and Readback | ||
____________ mA | ||
Iout | ____________ mA | |
29.945 A | ____________ A | |
Iout | ____________ mA | |
|
|
+80 mV Vout +120 mV
80.112 V
Vout +160 mV
+25 mA
Iout +35 mA
30.055 A
Iout +65 mA
2.2μV
2.2μV 1 mV 1 mV
21μA
21μA
15 mA
15 mA
* Enter your test results in this column.