Command Syntax

[SOURce]:VOLTage[:LEVel][:IMMediate][AMPLitude] <NRf+>

 

[SOURce][:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf+>

Parameters

Table 7-1

 

 

 

Default Suffix

V

 

 

 

*RST Value

Table 7-1

 

 

 

Examples

VOLT 200 MA

VOLTAGE:LEVEL 200 MA

 

VOLTAGE:LEVEL:IMMEDIATE:AMPLITUDE 2.5

 

VOLT:TRIG 20

VOLTAGE:LEVEL:TRIGGERED 20

Query Syntax

[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?

 

[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?

MAX

[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? MIN

 

[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]?

 

[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]? MAX

 

[SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]? MIN

Returned Parameters

<NR3> VOLT? and VOLT:TRIG? return presently programmed

 

immediate and triggered levels. If not triggered level is programmed,

 

both returned values are the same.

 

VOLT? MAX and VOLT? MIN return the maximum and

 

minimum programmable immediate voltage levels.

 

VOLT:TRIG? MAX and VOLT:TRIG? MIN return the

 

maximum and minimum programmable triggered voltage levels.

Related Commands

For VOLT

*SAV

*RCL

*RST

 

For VOLT:TRIG

ABOR

VOLT *RST

VOLT:PROT

This command sets the overvoltage protection (OVP) level of the power supply. If the output voltage exceeds the OVP level, then the power supply output is disabled and the Questionable Condition status register OV bit is set (see "Chapter 8 - Status Reporting"). An overvoltage condition can be cleared with the OUTP:PROT:CLE command after the condition that caused the OVP trip is removed. The OVP always trips with zero delay and is unaffected by the OUTP:PROT:DEL command.

Command Syntax

[SOURce]:VOLTage:PROTection[:LEVel] <NRf+>

*Alternate Syntax

[SOURce]:VOLTage:PROTection:AMPLitude <NRf+>

Parameters

Table 7-1

 

Default Suffix

V

 

*RST Value

MAX

 

Examples

VOLT:PROT 21.5

VOLT:PROT:LEV MAX

 

VOLTAGE:PROTECTION:LEVEL 145E-1

Query Syntax

[SOURce]:VOLTage:PROTection[:LEVel]?

 

[SOURce]:VOLTage:PROTection [:LEVel]? MIN

 

[SOURce]:VOLTage:PROTection [:LEVel]? MAX

Returned Parameters

<NR3> VOLT:PROT? returns presently programmed OVP level.

 

VOLT:PROT? MAX and VOLT:PROT? MIN return the

 

maximum and minimum programmable OVP levels.

Related Commands

OUTP:PROT:CLE

*RST *SAV *RCL

* Available to accommodate earlier power supply programs.

84 Language Dictionary