Performing The Tests

General Measurement Techniques

Figure C-1 shows the setup for the tests. Be certain to use load leads of sufficient wire gauge to carry the output current (see Table 8-1). To avoid noise pickup, use coaxial cable or shielded pairs for the test leads.

Programming the Power supply

Appendix A lists the programming voltage and current ranges. Enter the appropriate values from the front panel. The programming procedures assume you know how to operate the power supply from the front panel (see "Chapter 5 - Front Panel Operation").

Order of Tests

Perform the following tests for operation verification. Test 1 must be done first, followed by Tests 2 and 3 in any order.

1Turn-on Checkout

2Voltage Programming and Readback Accuracy

3Current Programming and Readback Accuracy

Turn-on Checkout

Perform the Turn-on Checkout as directed in "Chapter 3 - Turn-on Checkout".

Note The power supply must pass turn-on selftest before you can proceed with these tests.

Voltage Programming and Readback Accuracy

This test verifies that the voltage programming, GPIB readback, and front panel display functions are within specifications. Figure C-1(1) shows the setup for the tests. Measure the dc output voltage directly at the sense connections of the output terminals or bus bars. Connect the output as shown.

Table C-2. Voltage Programming and Readback Accuracy Tests

1

2

3

4

5

Action

Turn off the power supply and connect a DVM across the sense terminals (see Figure C-1(1)).

Turn on the power supply with no load and program the output for 0 volts and maximum programmable current.

Record voltage readings at DVM and on front panel display. (Subtract or add the specified readback limit to the actual output values).

Program voltage to full scale.

Record voltage readings of DVM and on front panel display. (Subtract or add the specified readback limit to the actual output values)

Normal Result

CV annunciator on. Output current near 0.

Readings within Low Voltage limits (see applicable test table).

Readings within High Voltage limits (see applicable test table).