ARPS Command1
VSET x
VSET xV
VSET xMV
ISET x
ISET xA
ISET xMA
VSET?
ISET?
VOLT?
IOUT?
OVP x
OVP xV
OVP xMV
OVP?
VMAX x VMAX xV
VMAX xMV
Table G-1. ARPS Commands
Description
These commands program output voltage. See Table
Initial condition: 0 V
These commands program output current. See Table
Initial condition: 0 A
These commands read voltage or current settings.
These commands measure and read output voltage or current.
NOTE: OVP commands do not work with Agilent 603xA supplies. These commands program the overvoltage protection. The OVP setting is programmed in either volts or millivolts. See Table
This command reads the OVP setting.
These commands program an upper limit (soft limit) to the voltage programming value that the power supply will accept. The programming ranges are the same as those used for VSET.
Similar SCPI
Command
VOLT
CURR
VOLT?
CURR?
MEAS:VOLT?
MEAS:CURR?
VOLT:PROT
VOLT:PROT?
(None)
IMAX x IMAX xA
IMAX xMA
These commands program an upper limit (soft limit) to the current (None) programming value that the power supply will accept. The programming ranges are the same as those used for ISET.
VMAX?
IMAX?
DLY x DLY xS DLY xMS
DLY?
OUT OFF OUT 0 OUT ON OUT 1
These commands read the soft voltage or current limits.
These commands program the delay time before a new output voltage or current is implemented or an RST, OUT ON, or CLR command is received. During the delay the CV, CC, and CR conditions cannot be reported as faults, and foldback protection is disabled.
This command reads the delay time setting.
These commands enable or disable the power supply output. The disabled state programs the output to relatively low voltage and current values. The unit will be able to implement commands even while the output is disabled.
Initial condition: OUT ON
(None)
OUTP:PROT:DEL
OUTP:PROT:DEL?
OUTP:STAT OFF OUTP:STAT 0 OUTP:STAT ON OUTP:STAT 1
FOLD OFF FOLD CC FOLD 2 FOLD 0 FOLD CV FOLD 1 FOLD CC FOLD 2
These commands enable or disable Foldback protection. Foldback protection disables the power supply output if the power supply switches to whichever mode (CV or CC) is defined as the fold (error) condition. Note that foldback protection is disabled during the DELAY period.
Initial condition: FOLD OFF
(None)
FOLD? |
| This command reads the Foldback setting. |
| (None) |
| 1 x = any | digit (within range) MA = milliampere MV = millivolt | MS = millisecond. | |
|
|
|
|
|