A
B
1.
2.
3.
From 1 to 16 direct supplies may be connected to 1 controller GPIB interface. Tighten connector thumbscrews by hand. Do not use a screwdriver.
Do not stack more than 3 connectors on a GPIB receptacle. GPIB cable (see Accessories in Chapter 1)
From 1 to 15 linked supplies may be connected to 1 direct unit. Either receptacle (Jl or J2) may be used as an input or an output.
Serial Link Cable (see Accessories in Chapter 1), 2 meters. 1 is supplied.
Maximum total length of all GPIB cables (including controller) not to exceed 20 meters. Use caution with individual lengths over 4 meters.
Maximum total length of all serial cables not to exceed 30 meters.
NOTES:
A direct power supply is connected to the controller interface and must have a unique primary GPIB bus address. The
The linked configuration uses 1 or more linked power supplies connected to each direct unit. Each linked unit has a unique secondary GPIB bus address and derives its primary address from the direct unit.
Figure 4-10. Controller Connections
36 User Connections