Initiate Subsystem
This subsystem enables the trigger system.
INIT
INIT:CONT
When a trigger is enabled with this command, an event on a selected trigger source causes the specified trigging action to occur. If the trigger subsystem is not enabled, all trigger commands are ignored. If INIT:CONT is OFF, then INIT enables the trigger subsystem only for a single trigger action. The subsystem must be enabled prior to each subsequent trigger action. If INIT:CONT is ON, then the trigger subsystem is continuously enabled and INIT is redundant.
| Command Syntax | INITiate[:IMMediate] | ||
| 
 | INITiate:CONTinuous <bool> | ||
| Parameters | For INIT[:IMM] | (None) | |
| 
 | For INIT:CONT | 01OFFON | |
| *RST Value | OFF | 
 | 
 | 
| Examples | INIT INITIATE:IMMEDIATE | ||
| 
 | INIT:CONT 1 | INITIATE:CONTINUOUS 1 | |
| Query Syntax | For INIT[:IMM] | (None) | |
| 
 | For INIT:CONT | INIT:CONT? | |
| Returned Parameters | <NR1> | 01 | 
 | 
| Related Commands | ABOR <GET> | *RST TRIG *TRG | |
Measure Subsystem
This subsystem returns output measurements.
MEAS:CURR?
MEAS:VOLT?
These queries return the voltage and current measured at the power supply’s sense terminals.
| Query Syntax | MEASure:CURRent[:DC]? | MEASure:VOLTage[:DC]? | |
| Parameters | (None) | 
 | 
 | 
| Default Suffix | A for MEAS:CURR? | V for MEAS:VOLT? | |
| Examples | MEAS:CURR? MEAS:VOLT? | MEASURE:VOLTAGE:DC? MV | |
| Returned Parameters | <NR3> | 
 | 
 | 
76 Language Dictionary