HP ColorFlex 13-1 manual Conditions of Illumination and Viewing, Power Required

Models: ColorFlex 13-1

1 24
Download 24 pages 5.13 Kb
Page 19
Image 19

Baud rate from 300 to 57600 No parity

8 data bits

1 stop bit.

Your output device must be configured in the same way in order to communicate with ColorFlex. Note that Universal Software’s maximum baud rate is 19200.

Note: Several baud rates are available with ColorFlex. Be sure to change the INSTRUMENT SETUP of ColorFlex to agree with the baud rate selected for your output device.

Handshake

Output from the ColorFlex serial port is controlled by XON/XOFF handshaking. Data output is stopped when the output device sends an XOFF character to ColorFlex. Data flow begins again when ColorFlex receives an XON character.

Conditions of Illumination and Viewing

Source

Dual-beam xenon flash lamp, >500,000 flashes

Source UV Content

Nominal match to D65

Integrating Sphere

63.5 mm diameter, high-efficiency white coating

Port Diameters/View Diameters

45/0 Model

31.8 mm (1.25”)/25.4 mm (1.0”)

Diffuse/8°

14.3 mm (0.56”)/8.0 mm (0.31”)

Power Required

External 120/220 V locking AC adapter

Power Cord Wire Color Code:

 

Color

Definition

220V Cord

Brown

LINE

 

Blue

NEUTRAL

 

Green/Yellow SAFETY

110V Cord

Black

LINE

 

White

NEUTRAL

 

Green

SAFETY

Instrument-Computer Ground Potential Check:

Perform this check if the power cord wiring is being changed, such as for replacement of the cord or if changing the plug.

07/02

ColorFlex

13-19

Page 19
Image 19
HP ColorFlex 13-1 manual Conditions of Illumination and Viewing, Power Required