SECTION 2—DISINFECTING THE TUB

SECTION 2—DISINFECTING THE TUB

WARNING

Disinfect the tub after EACH use to avoid resident infection and contamination of the tub. DO NOT disinfect the tub when it is occupied otherwise injury may occur.

Read and understand all information on Invacare brand disinfectant before use.

ALWAYS wear rubber gloves, an apron and a face shield when using disinfectant. In case of eye or skin contact, follow procedures on disinfectant container.

The disinfectant dispensing system has been factory tested and calibrated for use with Invacare disinfectant. The use of any disinfectant with a dilution ratio (½ oz. per gallon) or viscosity level different than that of Invacare disinfectant will result in an improper mixture of disinfectant and water. This may cause your system to be improperly disinfected - or in some instances, not disinfected at all. If using any disinfectant with a dilution ratio or viscosity level different than that of Invacare disinfectant or if switching from one type of disinfectant to another type, the disinfectant flowmeter MUST be calibrated before performing this procedure. Refer to Calibrating the Disinfectant Flowmeter on page 28.

DO NOT use the disinfectant wand for showering. The disinfectant wand is to be used for disinfecting ONLY.

CAUTION

NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice stone. The use of these types of cleaners will make the gel-coat finish of your tub scratched and dull. NEVER use cleaners containing iodine, bromine, Betadine or methylene blue. The use of these types of cleaners will cause the tub to stain. Never use cleaners containing bleach. Use of these types of cleaners will dry out the rubber seals and gaskets and the tub will not function properly.

Disinfecting the Tub Using the Closed Loop System

NOTE: For this procedure, refer to FIGURE 2.1.

NOTE: Ensure that the whirlpool jets and whirlpool inlet fitting are closed before starting this procedure.

1.Clean the entire tub surface to remove visible tissue residue and fluid.

2.Depending on the type of whirlpool jets that are in your tub, close all jets by performing one of the following:

Turn the center of the jet clockwise to a full stop.

Push the center of the jet in.

3.Close the whirlpool inlet by gently pulling the plug in the center of the inlet cover outward until it stops.

4.Open aerator by rotating aerator knob two turns counter‐clockwise.

5.Open disinfectant cabinet on the top of the tub.

6.Ensure that disinfectant siphon tube is placed in disinfectant container.

7.Hold the disinfectant wand over the interior of the tub.

8.Adjust water temperature to 105°F (40°C).

9.Turn disinfectant On/Off valve to the On position. NOTE: Liquid will begin flowing from the disinfectant wand into the tub.

10.Turn the adjustment knob on the flow meter until the floating ball rises to a level of 35 cc/min.

NOTE: At 35 cc/min., disinfectant is being mixed with water at the proper ratio.

11.Continue to hold the disinfectant wand over the interior of the tub.

NOTE: When disinfectant has totally filled the interior plumbing, liquid will begin to flow from the disinfectant wand.

NOTE: Invacare recommends that you thoroughly scrub all interior surfaces of the tub with disinfectant. The use of a long‐ handled brush will make this operation easier.

12.Spray the interior surfaces of the tub, including the whirlpool jets, overflow fitting and whirlpool inlet fitting with disinfectant. Scrub all interior surfaces.

13.Turn the disinfectant On/Off valve to the Off position.

14.Return disinfectant wand to its holder.

15.After scrubbing the interior surfaces, leave the disinfectant in the loop and on the tub surfaces for ten minutes.

16.Use the shower/massage wand on the pulse setting and direct the spray into both whirlpool jets to back flush until the water discharging from the whirlpool inlet is clear.

17.Rinse all interior tub surfaces.

18.Reverse STEPS 2 and 3 to open the whirlpool jets and whirlpool inlet.

DETAIL “A” -

Whirlpool Jet

WHIRLPOOL JET

 

AND INLET DETAILS

 

Turn Clockwise

Push In To

to Close

Close

Whirlpool Inlet

Plug

 

DETAIL “B” -

 

TUB DETAILS

Adjustment Knob

Flow meter

Ball

 

Disinfectant

Disinfectant Cabinet

Siphon Tube

Disinfectant

 

Disinfectant

On/Off Valve

 

Wand

 

 

Overflow

 

Whirlpool Jets

 

Whirlpool Inlet

FIGURE 2.1 Disinfecting the Tub Using the Closed Loop

System

Supine Bath Tub

12

Part No 1096926

Page 12
Image 12
Invacare IH6300ADH manual Disinfecting the TUB, Disinfecting the Tub Using the Closed Loop System

IH6300, IH6300ADH specifications

The Invacare IH6300ADH and IH6300 are advanced power wheelchairs designed to enhance mobility and independence for individuals with limited mobility. Featuring cutting-edge technology, these chairs are engineered to provide a safe, comfortable, and reliable mode of transportation for users.

One of the standout features of the Invacare IH6300 series is its robust and durable design. Built with a sturdy steel frame, these chairs can support a higher weight capacity, ensuring that they can accommodate a wide range of users. The ergonomic seating system provides optimal support and comfort during use, with adjustable seat depth and height options that cater to individual needs.

The IH6300ADH model takes comfort a step further with its enhanced cushioning and sophisticated suspension system, which absorbs shocks from uneven surfaces, allowing for a smoother ride. This is particularly beneficial for users who may travel over varied terrains or who may spend extended periods in their wheelchair.

Power and maneuverability are also key aspects of the Invacare IH6300 series. Both models come equipped with powerful motors that allow for swift movement and the ability to navigate inclines and tight spaces with ease. The intuitive joystick control enables users to operate the wheelchair effortlessly, offering precision in direction and speed.

Safety is paramount in the design of the IH6300 series. These wheelchairs feature intelligent braking systems that halt the chair when the joystick is released, preventing unintended movement. Additionally, the chairs are designed with anti-tip technology to enhance stability and prevent accidental falls.

The Invacare IH6300 series also boasts impressive battery life and efficiency, providing users with the freedom to travel longer distances without frequent recharging. The batteries are easily accessible for maintenance or replacement, ensuring ongoing reliability.

In summary, the Invacare IH6300ADH and IH6300 power wheelchairs stand out due to their exceptional build quality, user-friendly controls, and comprehensive safety features. They are ideal for individuals seeking mobility solutions that prioritize comfort, durability, and independence. With a range of options tailored to meet diverse user needs, the Invacare IH6300 series continues to be a leading choice in the realm of powered mobility devices.