Initial Startup of the Concentrator

SECTION 5—OPERATING INSTRUCTIONS

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.

 

 

 

LABEL

O2 PURITY

INDICATOR LIGHTS (LED)

SYMBOL

 

 

O2

SYSTEM OKAY

GREEN Indicator Light

O2 over 85% (± 2%)

 

 

 

 

 

O2 Between 73% (± 3%)

YELLOW Indicator light

 

to 85% (±2%)

A. YELLOW Solid

 

 

B. YELLOW Flashing Sensor

 

 

Failure

 

 

Call a qualified technician. Have a

 

 

back-up supply of oxygen ready.

 

 

 

 

SYSTEM FAILURE

RED Indicator Light

 

O2 Below 73% (±3%)

Continuous Audible Alarm

 

 

Sieve-GARD Compressor Shut-

 

 

down. Call a qualified technician.

 

 

Switch to a back-up supply of oxy-

 

 

gen immediately.

 

 

 

Front Panel Indicators - Units without SensO2

NOTE: For this procedure, refer to FIGURE 5.7.

Auto Shut Down

 

RED

Use Backup

 

 

 

 

 

Call Supplier

Normal

I/O

GREEN

FIGURE 5.7 Front Panel Indicators - Units without SensO2

Part No 1143482

21

Perfecto Series

 

 

2

Page 21
Image 21
Invacare IRC5PO2V Front Panel Indicators - Units without SensO2, Initial Startup of the Concentrator, O2 PURITY, Symbol