SECTION 7—TROUBLESHOOTING GUIDE
SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM: | PROBABLE CAUSE: | SOLUTION: | |
|
|
| |
Alarm: | Main Power Loss: | 1. Insert plug into outlet. | |
Short beeps, long | 1. Power cord not |
| |
pause | plugged in. |
| |
|
|
| |
Concentrator not | 2. No power at outlet. | 2. Inspect house circuit | |
operating, |
| breakers or fuses. If problem | |
| recurs, use a different outlet. | ||
power switch On. |
| ||
|
| ||
3. Tripped circuit | 3. Call Service Provider. | ||
Beep......Beep | breaker. |
| |
|
| ||
|
|
| |
Alarm: | System Failure: | 1a. Remove and clean cabinet | |
Continuous | 1. Unit overheating due | filters. | |
| to blocked air intake. |
| |
Concentrator not | 1b. Move oxygen | ||
| |||
| concentrator at least three | ||
operating, Power |
| ||
| inches away from walls, | ||
Switch On. |
| ||
| draperies or furniture. | ||
|
| ||
|
|
| |
Beeeeeeeep.... | 2. Insufficient power at | 2. DO NOT use extension | |
outlet. | cords. Move to another | ||
| |||
|
| electrical outlet or circuit. | |
|
|
| |
| 3. Internal repairs | 3. Call Service Provider. | |
| required. |
| |
|
|
| |
Alarm: | Low Flow Alarm: | 1. Inspect for kinks or | |
Rapid | 1. Kinked or blocked | blockages. Correct, clean or | |
| tubing, cannula or | replace item. Once | |
Beep....Beep... | humidifier. | corrected, turn power Off for | |
Beep...Beep |
| 60 seconds and then turn | |
|
| power back ON. | |
|
|
| |
| 2. Flowmeter set at 0.5 | 2. Reset flowmeter to | |
| L/min or less. | prescribed flowrate. If | |
|
| prescribed flowrate is below | |
|
| 0.5 L/min, use the Pediatric | |
|
| Flowmeter IRCPF16 (Only | |
|
| with Perfecto2 5). | |
|
|
|
Perfecto ™Series | 26 | Part No 1143482 |