SECTION 7—TROUBLESHOOTING GUIDE

SECTION 7—TROUBLESHOOTING GUIDE

SECTION 7—TROUBLESHOOTING GUIDE

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

 

 

 

Alarm:

Main Power Loss:

1. Insert plug into outlet.

Short beeps, long

1. Power cord not

 

pause

plugged in.

 

 

 

 

Concentrator not

2. No power at outlet.

2. Inspect house circuit

operating,

 

breakers or fuses. If problem

 

recurs, use a different outlet.

power switch On.

 

 

 

3. Tripped circuit

3. Call Service Provider.

Beep......Beep

breaker.

 

 

 

 

 

 

Alarm:

System Failure:

1a. Remove and clean cabinet

Continuous

1. Unit overheating due

filters.

 

to blocked air intake.

 

Concentrator not

1b. Move oxygen

 

 

concentrator at least three

operating, Power

 

 

inches away from walls,

Switch On.

 

 

draperies or furniture.

 

 

 

 

 

Beeeeeeeep....

2. Insufficient power at

2. DO NOT use extension

outlet.

cords. Move to another

 

 

 

electrical outlet or circuit.

 

 

 

 

3. Internal repairs

3. Call Service Provider.

 

required.

 

 

 

 

Alarm:

Low Flow Alarm:

1. Inspect for kinks or

Rapid

1. Kinked or blocked

blockages. Correct, clean or

 

tubing, cannula or

replace item. Once

Beep....Beep...

humidifier.

corrected, turn power Off for

Beep...Beep

 

60 seconds and then turn

 

 

power back ON.

 

 

 

 

2. Flowmeter set at 0.5

2. Reset flowmeter to

 

L/min or less.

prescribed flowrate. If

 

 

prescribed flowrate is below

 

 

0.5 L/min, use the Pediatric

 

 

Flowmeter IRCPF16 (Only

 

 

with Perfecto2 5).

 

 

 

Perfecto Series

26

Part No 1143482

2

 

 

Page 26
Image 26