SECTION 7—TROUBLESHOOTING GUIDE

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

 

 

 

YELLOW or RED

1. Low oxygen purity.*

1. Clean or Replace filters.

light Illuminated.

 

 

2. Kinked or blocked

2. Inspect for kinks or

 

Alarm:

tubing, cannula or

blockages. Correct, clean or

humidifier.*

replace item. Once

Continuous

 

corrected, turn power Off for

On RED light

 

 

60 seconds and then turn

only.

 

 

power back On.

 

 

 

 

 

* Only applies to

3. Flowmeter set at 0.5

3. Check flowmeter is set to

L/min or less.*

1.0 L/min or more.

IRC5PO2.

 

 

 

4. Unit overheating due

4a. Remove and clean cabinet

 

to blocked air intake.

filters.

 

 

 

 

 

4b. Move concentrator at

 

 

least three inches from walls,

 

 

draperies, and furniture.

 

 

 

 

5. Insufficient power at

5. DO NOT use extension

 

outlet.

cords. Move to another

 

 

electrical outlet or circuit.

 

 

 

 

6. Internal repairs

6. Call Service Provider.

 

required.

 

 

 

 

GREEN light with

1. Internal repairs

1. Call Service Provider.

YELLOW Light

required.

 

Flashing.

 

 

 

 

 

Part No 1143482

27

Perfecto Series

 

 

2

Page 27
Image 27
Invacare IRC5PO2V, IRC5PO2W user manual Troubleshootingguide, Symptom, Solution, Probable Cause, Alarm