PREVENTIVE MAINTENANCE RECORD

Part No 1143482

25

Perfecto2™Series

FIGURE 6.2 Preventive Maintenance Record

PREVENTIVE MAINTENANCE RECORD

ON EACH INSPECTION

Record Date of Service

Record Elapsed Hours On Hour Meter

Clean Cabinet Filter(s)

Check Prescribed L/min. Flow Rate

EVERY 4,380 HOURS

Check Oxygen Concentration

(UNITS WITHOUT SensO 2 )

EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST

Check Oxygen Concentration

(UNITS WITH SensO 2 )

DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS

Clean/Replace Cabinet Filter(s)

Check Outlet HEPA Filter*

Check Compressor Inlet Filter*

Check Power Loss Alarm

AS REQUIRED

Clean Heat Exchanger

Replace Exhaust Muffler

Rebuild Top End of Compressor

*NOTE: Refer to Preventive Maintenance section of Service Manual. NOTE:

4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months. 26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.

MODEL NO. I RC

SERIAL NO.

SECTION 6—MAINTENANCE

Page 25
Image 25
Invacare IRC5PO2V Preventive Maintenance Record, On Each Inspection, Model No. I Rc Serial No, EVERY 4,380 HOURS