SECTION 4—TYPICAL PRODUCT PARAMETERS

SECTION 4—TYPICAL PRODUCT PARAMETERS

SECTION 4—TYPICAL PRODUCT PARAMETERS

Electrical Requirements:

120 VAC ± 10% (132 VAC/108 VAC), 60 Hz

 

 

Rated Current Input:

3 A

 

 

Sound Level:

43 ± 2 dBA Average for Perfecto2 Models and

 

Perfecto2V Model

 

40 ± 2 dBA Average for Perfecto2W Models

 

 

Altitude:

All Perfecto2 Models: Up to 8,000 ft (2438

 

meters) above sea level without degredation of

 

concentration levels. From 8,000 ft to 13,129

 

(4000 meters) below 90% efficiency.

 

All Perfecto2 Models while filling a

 

HomeFill System: Up to 6,000 ft (1828 meters)

 

above sea level without degradation of

 

concentration levels. From 6,000 ft (1828 meters)

 

to 13,129 ft (4000 meters) below 90% efficiency.

 

 

*Oxygen Output

87% to 95.6% for flows ranging at 0.5 to 5 L/min

Concentration Levels:

 

 

 

Maximum Outlet Pres-

5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)

sure:

 

 

 

Flow Range:

0.5 to 5 L/min

 

For flowrates less than 1 L/min, we recommend

 

the use of the Invacare Pediatric Flowmeter

 

IRCPF16 Accessory.

 

 

Low Flow Alarm:

0 L/min to 0.5 L/min

 

Rapid Audible Alarm Beeping (No Accessories

 

Connected).

 

 

Average Power

Perfecto2 and Perfecto2W Models: 280 W

Consumption:

(Operating @ 3 L/min)

 

Perfecto2V Model: 325 W

 

 

Pressure Relief Mecha-

35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)

nism Operational at:

 

 

 

Perfecto Series

12

Part No 1143482