Invacare Model, 3752G user manual Operating Instructions, Detail “A”, Detail “B”, Detail “C”

Models: 3752G Model

1 36
Download 36 pages 34.9 Kb
Page 22
Image 22
Detail “A”

5 OPERATING INSTRUCTIONS

1.Use the drain plug to close the drain.

2.Remove and disassemble all jet assemblies. Lay all pieces in the footwell of the tub.

Rotate each jet assembly counterclockwise and pull away from the jet casing. (Detail “A”).

Remove each impeller cap by rotating it clockwise (Detail “B”).

Remove each impeller (Detail “C”).

Detail “A”

Detail “B”

Jet Casing

Impeller Cap Detail “B”Detail “C”

Impeller

Jet Assembly

Detail “C”

Impeller Manual background

3.Remove the waist straps from the mounting attachment points. Put the straps into the footwell of the tub.

4.Clean the pieces and spray all surfaces of the tub with Dispatch® Cleaner and Disinfectant. Take a long handled brush and thoroughly clean the interior surfaces of the tub and the jet casings.

5.Allow the Dispatch Cleaner and Disinfectant to sit on the surfaces for one minute.

6.Rinse all surfaces and pieces in the footwell of the tub with water.

7.Use a clean towel to dry all tub surfaces and all pieces in the footwell of the tub. Allow the waist straps to air dry.

￿DO NOT put the waist straps in the dryer, otherwise damage to the straps may occur.WARNING

Invacare® TheraPure™ Recessed Whirlpool Tubs

22

Part No. 1167458

Page 22
Image 22
Invacare Model Operating Instructions, Detail “A”, Detail “B”, Detail “C”, Invacare TheraPure Recessed Whirlpool Tubs