MAINTENANCE/TROUBLESHOOTING

B. CABINET

WARNING

UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet.

1.Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.

C. HUMIDIFIER

NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these steps:

1.Clean the humidifier every day.

2.Wash it in soapy water and rinse it with a solution of ten (10) parts water and one (1) part vinegar.

3.Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier.

WARNING

DO NOT overfill Humidifier Bottle.

CONCENTRATOR

TROUBLESHOOTING GUIDE

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

 

 

 

 

 

Alarm:

Main Power Loss:

1.

Insert plug into outlet.

SHORT BEEPS,

1. Power cord NOT

2.

Inspect house circuit

LONG PAUSE

 

plugged In.

 

 

breakers or fuses. If

 

 

 

 

Concentrator

2.

No power at outlet.

 

problem recurs, use a

NOT operating,

3.

Tripped circuit breaker.

 

different outlet.

power switch ON.

3.

Call Service Provider.

Beep......Beep

 

 

 

 

 

 

 

 

 

 

Alarm:

System Failure:

 

 

CONTINUOUS

1.

Unit overheating due

1a. Remove and clean

Concentrator

 

to blocked air intake.

 

cabinet filters.

 

 

1b.

Move oxygen concentrator at

NOT operating,

 

 

Power Switch ON.

 

 

 

least 7.5 cm (3 in) away from

 

 

 

walls, draperies or furniture.

Beeeeeeeep....

 

 

 

 

 

 

 

 

2.

Insufficient power at

2.

Do not use extension

 

 

outlet.

 

cords. Move to another

 

 

 

 

electrical outlet or circuit.

 

3.

Internal repairs required.

3.

CallServiceProvider.

 

 

 

 

 

 

 

 

 

 