TROUBLESHOOTING

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

YELLOW or RED

1. Low oxygen purity.*

1.

Clean or replace filters.

LIGHT

2. Kinked or blocked

2.

Inspect for kinks or

Illuminated.

tubing, cannula or

 

blockages. Correct,

 

 

Alarm:

humidifier.*

 

clean or replace item.

 

 

 

Once corrected, turn

CONTINUOUS

 

 

 

power OFF for 60

On red light only.

 

 

 

seconds and then turn

 

 

 

 

power back ON.

 

3. Flowmeter set at

3.

Check flowmeter is set

 

1/2 L/min. or less.*

 

to 1.0 L/min. or more.

 

 

 

 

Refer to Recommended

 

 

 

 

End-User Guidelines in

 

 

 

 

the Operating Instructions

 

 

 

 

of this manual.

 

4. Unit overheating

4a.

Remove and clean

 

due to blocked air

 

cabinet filters.

* Only applies to

intake.

4b.

Move concentrator at

SensO2 models.

 

 

 

least 7.5 cm (3 in) from

 

 

 

 

walls, draperies, and

 

 

 

 

furniture.

 

5. Insufficient power at

5.

Do not use extension

 

outlet.

 

cords. Move to another

 

 

 

 

electrical outlet or circuit.

 

6. Internal repairs required.

6.

Call Service

 

 

 

 

Provider.

 

 

 

 

 

GREEN LIGHT

1.

Internal repairs required.

1.

Call Service Provider.

with YELLOW

Light Flashing.

 

 

 

 

 

 

 

 

Alarm:

LOW FLOW ALARM

1a.

Inspect for kinks or

RAPID

1a.

Kinked or blocked

 

blockages. Correct,

Beep....Beep...

 

tubing, cannula or

 

clean or replace item.

 

humidifier.

 

Once corrected, turn

Beep...Beep

 

 

 

 

 

power OFF for 60

 

 

 

 

 

 

 

 

seconds and then turn

 

 

 

 

power back ON.

 

1b. Flowmeter set at

1b.

Reset flowmeter to

 

 

0.5 L/min. or less.

 

prescribed flowrate.

 

 

 

 

Flowrates less than

 

 

 

 

1L/min. are not

 

 

 

 

recommended.

 

 

 

 

 

21

Page 21
Image 21
Invacare PLATINUM 5 manual Troubleshooting, Light