TROUBLESHOOTING
SYMPTOM: | PROBABLE CAUSE: | SOLUTION: | |||
YELLOW or RED | 1. Low oxygen purity.* | 1. | Clean or replace filters. | ||
LIGHT | 2. Kinked or blocked | 2. | Inspect for kinks or | ||
Illuminated. | |||||
tubing, cannula or |
| blockages. Correct, | |||
|
| ||||
Alarm: | humidifier.* |
| clean or replace item. | ||
|
|
| Once corrected, turn | ||
CONTINUOUS |
|
|
| power OFF for 60 | |
On red light only. |
|
|
| seconds and then turn | |
|
|
|
| power back ON. | |
| 3. Flowmeter set at | 3. | Check flowmeter is set | ||
| 1/2 L/min. or less.* |
| to 1.0 L/min. or more. | ||
|
|
|
| Refer to Recommended | |
|
|
|
| ||
|
|
|
| the Operating Instructions | |
|
|
|
| of this manual. | |
| 4. Unit overheating | 4a. | Remove and clean | ||
| due to blocked air |
| cabinet filters. | ||
* Only applies to | intake. | 4b. | Move concentrator at | ||
SensO2 models. |
|
|
| least 7.5 cm (3 in) from | |
|
|
|
| walls, draperies, and | |
|
|
|
| furniture. | |
| 5. Insufficient power at | 5. | Do not use extension | ||
| outlet. |
| cords. Move to another | ||
|
|
|
| electrical outlet or circuit. | |
| 6. Internal repairs required. | 6. | Call Service | ||
|
|
|
| Provider. | |
|
|
|
|
| |
GREEN LIGHT | 1. | Internal repairs required. | 1. | Call Service Provider. | |
with YELLOW | |||||
Light Flashing. |
|
|
|
| |
|
|
|
| ||
Alarm: | LOW FLOW ALARM | 1a. | Inspect for kinks or | ||
RAPID | 1a. | Kinked or blocked |
| blockages. Correct, | |
Beep....Beep... |
| tubing, cannula or |
| clean or replace item. | |
| humidifier. |
| Once corrected, turn | ||
Beep...Beep |
|
| |||
|
|
| power OFF for 60 | ||
|
|
|
| ||
|
|
|
| seconds and then turn | |
|
|
|
| power back ON. | |
| 1b. Flowmeter set at | 1b. | Reset flowmeter to | ||
|
| 0.5 L/min. or less. |
| prescribed flowrate. | |
|
|
|
| Flowrates less than | |
|
|
|
| 1L/min. are not | |
|
|
|
| recommended. | |
|
|
|
|
|