SPECIFICATIONS
SPECIFICATIONS
Alternating Current
Type BF Equipment
Running
NOT Running
DO NOT Smoke
|
| Class II |
|
Electrical Requirements: | 230 VAC ± 10% (253 VAC/207 VAC), 50 Hz | ||
Rated Current Input: | 2.0 A | ||
Sound Level: | 44 dB Average | ||
Altitude: | Up to 1828 m(6,000 ft) above sea level without | ||
|
|
| degradation of concentration levels. NOT |
|
|
| RECOMMENDED FOR USE ABOVE |
|
|
| 1828 M (6000 FT). |
*Oxygen Output
Concentration Levels: ALL 5LXA/5LXO2A MODELS 93% minimum at 1 to 3 L/min 91% minimum at 4 L/min
87% minimum at 5 L/min (maximum recommended flow)
*Stated concentration levels achieved after initial
Maximum Outlet Pressure: 34.5 ± 3.45 kPa (5 ± 0.5 psi)
5LXO2 INDICATORS:
Label | Status | Indicator Lights (LED) |
Symbol |
|
|
|
|
|
| SYSTEM FAILURE | RED Continuous Audible Alarm |
| Below 73% | Compressor Shutdown |
| 73% to 85% | YELLOW |
O2 | SYSTEM OKAY | GREEN |
Over 85% |
| |
|
| |
|
|
|
|
|
|
9