SPECIFICATIONS

SPECIFICATIONS

Alternating Current

Type BF Equipment

Running

NOT Running

Attention-Consider Accompanying Documents

DO NOT Smoke

 

 

Class II

 

Electrical Requirements:

230 VAC ± 10% (253 VAC/207 VAC), 50 Hz

Rated Current Input:

2.0 A

Sound Level:

44 dB Average

Altitude:

Up to 1828 m(6,000 ft) above sea level without

 

 

 

degradation of concentration levels. NOT

 

 

 

RECOMMENDED FOR USE ABOVE

 

 

 

1828 M (6000 FT).

*Oxygen Output

Concentration Levels: ALL 5LXA/5LXO2A MODELS 93% minimum at 1 to 3 L/min 91% minimum at 4 L/min

87% minimum at 5 L/min (maximum recommended flow)

*Stated concentration levels achieved after initial warm-up period (approximately 30 minutes).

Maximum Outlet Pressure: 34.5 ± 3.45 kPa (5 ± 0.5 psi)

5LXO2 INDICATORS:

Label

Status

Indicator Lights (LED)

Symbol

 

 

 

 

 

 

SYSTEM FAILURE

RED Continuous Audible Alarm Sieve-GARDTM

 

Below 73%

Compressor Shutdown

 

73% to 85%

YELLOW

O2

SYSTEM OKAY

GREEN

Over 85%

 

 

 

 

 

 

 

 

 

9

Page 9
Image 9
Invacare PLATINUM 5 manual Specifications, 5LXO2 Indicators, Recommended for USE Above